Particle Sensor KS-18FX

High sensitivity allows measurement of 0.04 μm particles in liquid


Image

Categories

Measuring Instruments > Particle Counting Systems > Liquid Particle Counter
Semiconductor > Particle Counter Semiconductor > Liquid Particle Counter Semiconductor

Description

Detects particles down to 0.04 μm size, at a flow rate of 10 mL per minute
User selectable particle size range up to 10 channnels within measuring range using KE-40B1
Integrated leak sensor with alarm output
Sapphire flow cell

< Specification >

Optical system: Light-scattering method
Light source: Diode pumped solid state laser
(rated output 500 mW, wavelength 532 nm)
Laser product class: Class 1, IEC 60825-1
Light detector: Silicon photodiode
Materials of parts exposed to sample: Sapphire, PFA
Allowable sample fluid type:
Fluids which do not corrode the fluid contact materials
Size range: 4 channels (factory default)
≧0.04 μm, ≧0.08 μm, ≧0.1μm, ≧0.15 μm
User selectable channels: 1 to 10 channels (0.04 μm to 0.15 μm)
Flow rate: 10 mL/min
Counting efficiency: 3 %
Maximum particle number concentration:
30 000 particles/mL (coincidence loss 10 %)
Dimensions and weight:147 (H) x 272 (W) x 442 (D) mm
(excluding protruding parts), approx. 12 kg