Particle Sensor KS-42D

2 µm to 100 µm Measurement Range Allows Monitoring of Coarse Particles


Image

Categories

Measuring Instruments > Particle Counting Systems > Liquid Particle Counter
Semiconductor > Particle Counter Semiconductor > Liquid Particle Counter Semiconductor

Description

Detects particles down to 2 µm size, at a flow rate of 25 mL per minute
Particle size range (8channels, factory default setting)
≥2µm, ≥3µm, ≥5µm, ≥7µm, ≥10µm, ≥25µm, ≥50µm, ≥100µm
Integrated leak sensor with alarm output
User selectable channels within measurement range
(using KE-40B1 function)

< Specification >

Optical system: Light-obscuration method
Light source: Laser diode (wavelength 780 nm, rated output 5 mW)
Laser product classification: Class 1, IEC 60825-1
Fluid contact materials: Synthetic quartz, PFA, perfluoro
(fluorocarbon rubber)
Allowable sample fluid types: Fluids which do not corrode the fluid contact materials
Particle size range: 8 channels(factory default) :
≥2µm, ≥3µm, ≥5µm, ≥7µm, ≥10µm, ≥25µm, ≥50µm, ≥100µm (≥150µm support available as option)
User selectable channels: 1 to 10 channels, setting made from Controller
Setting range: 2 µm to 100 µm
Count efficiency: 100 ± 20 %
Sample flow rate: 25 mL/min
Maximum particle concentration: 10 000 particles/mL (coincidence loss 10 % for 10 µm particles)
Allowable sample fluid pressure: -80 to 300 kPa (gauge pressure)
Power: DC12 V (supplied from KE-40B1)
Ambient conditions for operation: 5 °C to 40 °C, less than 90 % RH (no condensation)
Dimensions and weight: 125(H) × 140 (W) × 150 (D) mm (excluding protruding parts), Approx. 2.2 kg

Need help?

Not sure exactly what you’re looking for or just want clarification? We’d be happy to chat with you and clear things up for you. Anytime!

Call anytime

+49 8131 998099

Email us

info@qc-quality-control.de

Address

QC-Quality Control GmbH
Dieselstrasse 11
D-85757 Karlsfeld
Phone: +49 8131 998099
Fax: +49 8131 998101
E-Mail: info@qc-quality-control.de
Contact